Radial Catheter Selection
A guide to catheter selection for transradial coronary angiography and intervention.
By Nicholas R. Balaji, MD; Pinak B. Shah, MD; and Frederic S. Resnic, MD, MSc
Radial artery access for cardiac catheterization/percutaneous coronary intervention (PCI) has recently gained significant attention due to the associated reduction in bleeding complications with radial compared to femoral access. The major limitation to widespread adoption of radial access is that coronary intubation from a transradial approach is more challenging, and as a result, the learning curve is steep. The widely used preformed catheter shapes are designed to easily find the coronary ostia when used from the femoral artery. When approaching the coronaries from the radial artery, additional considerations apply.
CATHETER SELECTION FOR DIAGNOSTIC CARDIAC CATHETERIZATION
Native Coronary Arteries
Given the widespread familiarity with Judkins catheters,
it is not surprising that these remain the most commonly
used shapes for coronary angiography, regardless of
access site.1 When using a Judkins approach, the first
obstacle to overcome is proximal innominate artery tortuosity
and accessing the ascending
aorta. Many operators will
lead with the Judkins right (JR)
catheter so that a 0.035-inch wire
can be directed into the aorta,
avoiding the vertebral and carotid
arteries. Once in the aorta, this
curve can also assist in accessing
the ascending limb, in some cases
aided by a deep breath from the
patient. In most cases, a JR 4 will
engage the right coronary artery
(RCA) using a standard femoral
technique, although in some
patients, upsizing to a JR 5 from
the right radial will allow for more
stable intubation. Leading with
this catheter also allows the operator to easily obtain a nonselective angiogram of the left
main (LM) ostium before engaging with a Judkins left (JL)
catheter (Figure 1A). This can be helpful when LM disease
is a concern, as occasionally the JL will need to be
engaged with the wire still in place. After exchanging
catheters, the tip of the JL will typically be in the right
coronary cusp. Gentle retraction will result in movement
of the catheter into the left cusp and often into
the LM ostium. When utilizing the right radial, a JL 3.5
will typically allow better engagement. Unfortunately,
the catheter tip may point up to the roof of the LM, and
gentle injection is imperative (Figure 1B).
Other femoral catheters, such as the Amplatz shapes, can be helpful when anatomic variations prove challenging for the Judkins shapes. With dilated aortic roots, the larger Amplatz left (AL) curve will often provide the best orientation for engagement of the LM. An anterior RCA is also best approached with an AL 0.75 or 1, as with femoral access. Although the benefits in terms of familiarity with these catheter shapes and engagement techniques are significant, the main disadvantage of using these catheters is the necessity for multiple catheter exchanges. A higher number of catheter exchanges increases vascular dysfunction and may provoke radial artery spasm, leading to patient discomfort and difficult catheter manipulation.2 Multiple exchanges can also prolong the procedure duration and increase radiation exposure to the patient and operator.
To avoid the disadvantages of a dual-catheter strategy, some operators prefer to use a single-catheter technique. In the United States, the devices that are most commonly used to engage both left and right coronaries are the Tiger (Terumo Interventional Systems, Somerset, NJ) and the Kimny (Boston Scientific Corporation, Natick, MA) catheters.1 These catheters were specifically designed to be used from the right radial approach and require a different engagement technique. The shapes of these catheters are similar, as is the technique for intubation of the coronaries (Figure 2A and 2B). When advanced over a guidewire into the ascending aorta, they may end up in either the left or right coronary cusp. In our experience, they typically fall into the right cusp, and engagement of the left cusp requires careful withdrawal until the tip retracts over the right aortic valve leaflet and moves into the left coronary cusp. Depending on the patient's anatomy, the catheter may engage from above with slight clockwise or counterclockwise rotation or, more commonly, may require advancement up the left cusp to engage from below with slight withdrawal at the end to achieve a coaxial position, similar to the typical Amplatz technique. To engage the right coronary, the catheter is then slightly withdrawn from the left, rotated just slightly over the valve, and advanced into the right cusp. Once in the right coronary cusp, slow withdrawal with clockwise rotation will engage similar to the Judkins technique.
Smaller studies have shown these catheters to be safe and effective and suggest that they may reduce fluoroscopy time and total procedure time without sacrificing the quality of angiography.3,4 However, anatomic variants are common, and just as with femoral catheterization, no single catheter will be successful in every case. It is important to recognize this early and switch to an alternative catheter shape, if needed.
Coronary Artery Bypass Grafts
Previous bypass has been shown to be an independent
predictor of transradial failure, but with sufficient
operator experience, most patients can undergo coronary
angiography via the radial approach after coronary
artery bypass graft surgery.5 The ipsilateral radial artery
is used if the mammary artery has been utilized as a
conduit. If both the right and left internal mammary arteries (IMAs) have been used, the contralateral IMA
can be engaged from the radial artery. In this situation,
the patient would be most comfortably approached
from the right radial. An IMA catheter will allow for
right IMA injection and can then be advanced into the
descending aorta over a 0.035-inch wire. The catheter
can then be oriented to allow passage of a 0.035-inch
hydrophilic wire into the left subclavian, axillary, and
brachial arteries. At this point, inflating the brachial
blood pressure cuff will provide enough support on the
wire to allow the IMA catheter to be advanced into the
subclavian artery, and the left IMA can be engaged in
the usual fashion.
Saphenous vein grafts (SVGs) from the aorta are typically best approached via the left radial, and most commonly, an AL or JR catheter can be used to reach all the grafts. In some cases, a multipurpose catheter may be helpful, specifically when engaging an inferiorly oriented graft to the RCA.
CATHETER SELECTION FOR PCI
The selection of a guide catheter for coronary interventions involves an additional level of consideration. The initial concerns regarding ease of coronary intubation, coaxial alignment, and the technique to achieve these are identical to those of diagnostic catheters. However, for guides, the importance of backup support to allow delivery of interventional equipment is paramount. Support is derived from the intrinsic stiffness of the guide, as well as contact against the opposite wall of the aorta.
Left Coronary Intervention
Common catheter shapes, such as the Judkins, XB
(Cordis Corporation, Bridgewater, NJ), and EBU
(Medtronic, Inc., Minneapolis, MN), provide significantly
more backup when used from a femoral approach.6
Despite this, the common femoral shapes are still the
most widely used guide catheters, with XB/EBU and JL
being most common for the left coronary.1 Downsizing
slightly to the JL 3.5 has been shown to increase backup
support in in vitro studies, but it is unclear if this is true
of the XB/EBU shape as well.6 The Ikari guide (Terumo
Interventional Systems) was specifically designed to provide
improved backup support from the right radial
approach. The Ikari shape was designed for left coronary
interventions and takes into account the angle between
the brachiocephalic and ascending aorta (Figure 2C).7 It
allows for backup support from the contralateral aortic
wall and coaxial engagement of the LM ostium. The
catheter will typically enter the left sinus spontaneously,
where withdrawal of the wire and occasional clockwise
or counterclockwise torque will result in LM engagement.8
Sizing of this catheter is similar to the JL, with an Ikari left
4 guide being appropriate if a JL 4 was an appropriately
sized diagnostic catheter.9
Right Coronary Intervention
For the right coronary, the JR is again the most frequently
used guide catheter, with the Amplatz right a distant
second.1 Unfortunately, lack of backup support
remains an issue, particularly with the JR catheter, which
does not contact the contralateral aortic wall at all. Some
operators favor “deep-seating” the guide, especially when
using a smaller 5-F guide, but there is some risk of dissection
with this technique. When additional backup support
is needed, a smaller AL 0.75 is useful, and the technique
used for engagement is similar to the approach
from the femoral artery. Alternatively, the Ikari right
catheter may be used (Figure 2D). This catheter's threedimensional
curve takes into account the angles between
the innominate and the right coronary ostium, allowing
coaxial intubation with backup support from the contralateral
aortic wall. When advanced over a 0.035-inch
wire, the catheter tip will typically end up in the left
coronary sinus. Withdrawing the wire but leaving it in
the guide prevents kinking of the catheter while it is gently
rotated into the right sinus and into the coronary
ostium.8 The MAC guide (Medtronic, Inc.) is another
long-tip guide that provides backup from the contralateral
aortic wall in a similar fashion. Although not commonly
used, either of these specialty catheters should be
considered when more familiar catheters do not provide
adequate support.
Bypass Graft Intervention
IMA interventions are generally performed using an
IMA guide from the ipsilateral radial artery. If the lesion
to be treated is in the distal IMA or in the native artery
beyond the anastomosis, consideration should be given
to using a 90-cm guide to ensure that the usable length
of balloon and stent catheters will be sufficient to reach
the lesion.
Guide support for SVG intervention can be especially difficult due to the location of the grafts and the proximity to the innominate or left subclavian artery. When approached from the right radial, a significant portion of the catheter's curve is located in the innominate artery and does not provide support along the wall of the aorta. For SVGs to left coronary artery vessels, a JR guide can be considered but often will not have a long enough tip to engage the graft and will not provide backup support from the contralateral aortic wall. An AL guide from the left radial provides the best support, but a coaxial position can be difficult to achieve.
For SVG interventions to RCA branches, a JR 4 guide can be considered but has similar limitations as when used for left-sided grafts. Our preference is to use a multipurpose guide, particularly for grafts with an inferior takeoff.
FUTURE DIRECTIONS
An important limitation to radial access is the risk of radial artery occlusion, which occurs in approximately 5% to 10% of patients at discharge, with 2% to 7% persistently occluded at 1-month follow-up.11,12 A major predictor of radial artery occlusion is a larger ratio of sheath diameter to radial artery diameter, and therefore, investigation is ongoing to improve sheath and guide technology so that PCI can be performed via a smaller profile.10,13 Although 5-F catheters may provide less backup support and more difficult angiographic visualization, they have the advantage that they can often be easily and safely advanced deep into the coronary artery, thereby providing excellent support for device delivery. Of course, many devices cannot be delivered through a 5-F guiding catheter.
Nearly all PCIs can be successfully completed with a 6-F system, including cases involving thrombectomy, rotational atherectomy, and bifurcation balloon angioplasty and stenting with the “Culotte” technique. The majority of adult patients will have radial arteries that can accommodate such equipment (Figure 3).10 However, there are cases in which the operator will need a larger guiding catheter either for enhanced support or additional space for multiple wires, catheters, and stents. Unfortunately, with the use of larger sheaths, the risk of radial artery occlusion increases.13 In an effort to maintain a smaller profile, several operators have described the use of the “sheathless” guide technique.14 Once radial access is achieved and a 0.035-inch wire is advanced into the ascending aorta, a 7-F guide can then be advanced over a 5-F, 125-cm diagnostic catheter (such as a multipurpose shape) that acts as a dilator (Figure 4). Once in the ascending aorta, the guide can be advanced over the inner dilator catheter and into the coronary artery. This technique allows for larger guide diameter with less radial trauma. For instance, a 7-F guide catheter has the same outer diameter as a 5-F sheath. Guide technology must continue to improve with the development of hydrophilic-coated guides and better dilator technology for sheathless PCI to more readily come into practice.
CONCLUSION
With the increasing use of radial artery access for PCI, it is important for operators to become familiar with currently available and emerging technologies. Appropriate application of techniques and technology to the various clinical scenarios and diverse patient populations will allow physicians to achieve optimal results with transradial intervention.
Nicholas R. Balaji, MD, is an interventional cardiology fellow at Brigham and Women's Hospital in Boston, Massachusetts. He has disclosed that he holds no financial interest in any product or manufacturer mentioned herein.
Pinak B. Shah, MD, is Director of the Interventional Cardiology Training Program at Brigham and Women's Hospital in Boston, Massachusetts. He has disclosed that he holds no financial interest in any product or manufacturer mentioned herein.
Frederic S. Resnic, MD, MSc, is Director of the Cardiac Catheterization Laboratory at Brigham and Women's Hospital in Boston, Massachusetts. He has disclosed that he holds no financial interest in any product or manufacturer mentioned herein. Dr. Resnic may be reached at (857) 307-1992; fresnic@partners.org.