Advertisement
Advertisement
November 2009
Imaging for ASD and PFO Closure
The requirements and future direction of imaging for percutaneous atrial septal defect and patent foramen ovale closure procedures.
Percutaneous atrial septal defect (ASD) and patent foramen ovale (PFO) closure procedures are now the most common congenital intervention performed in the cardiac catheterization laboratory. Although this procedure has used ultrasound imaging guidance since the early 1990s,1 the frequency of ultrasound imaging in the cardiac catheterization laboratory has increased greatly during the past decade. Simultaneously, with the emergence of new imaging modalities,2 many congenital interventional cardiologists, rather than their noninvasive colleagues are performing the imaging as part of the atrial septal device closure procedure.
BACKGROUND
Atrial septal shunt is a common problem in congenital
heart disease; 20% to 25% of the population has a PFO,3
and 30% to 40% of patients with a congenital heart problem
have an ASD.4 Since the first reports of transcatheter
device closure for ASD and PFO, further development of
transcatheter techniques and improved atrial septal occlusion
device design have resulted in this procedure becoming
the treatment of choice for this patient population.5 In
addition, imaging requirements during these procedures
has changed due to the ongoing evolution toward better
imaging for patients with congenital heart disease.
IMAGING REQUIREMENTS
Fluoroscopy as a Second Imaging Modality or Alone
The majority of patients with congenital heart disease
who enter the cardiac catheterization laboratory will have
fluoroscopic imaging as one tool to guide the interventional
procedure. Although the standard has been to use a
combination of fluoroscopy and echocardiographic imaging,
there is no agreement that the congenital invasive cardiologists
use a biplane laboratory or a single-plane laboratory
for atrial septal interventions (Figure 1). Although
there is no requirement for biplane capabilities when using
a single-plane laboratory, the invasive cardiologist will usually
have to maneuver the camera from the anteroposterior
view to the left anterior oblique view to assess the device
and outline the septum appropriately. In addition, the use
of fluoroscopy is of greater value with the Helex septal
occluder as compared to the Amplatzer septal occluder
(AGA Medical Corporation, Plymouth, MN). This is inherent
in the device design and the need to confirm the position
of the lock loop with the Helex device.
Fluoroscopy alone can be performed for some ASD and PFO device closure procedures. With this single-imaging technique, the congenital cardiologist can use angiography to outline the septal anatomy and determine which device to implant. Although the advantages of this technique are decreased cost and procedural duration, there may be an increase in adverse events, such as device malposition and residual leak. Consequently, although most congenital invasive cardiologists may consider this technique for small shunts, the majority of septal closure procedures require a combination of imaging modalities.
2D Transesophageal Echocardiography
With Fluoroscopy
Transesophageal echocardiography (TEE) has been the
mainstay for guidance (along with fluoroscopy) during atrial
septal interventions during the past few decades.6 The
TEE imaging modality (Figure 2) has advantages related to
the cost of the probe, multiplane imaging capabilities, and
avoidance of an additional large venous sheath. On the
other hand, TEE has disadvantages, including patients
requiring a general anesthetic, the cost and time utilization
for having an anesthesiologist and a second cardiologist
involved for the duration of the procedure, and patient discomfort
during the procedure, resulting in more sedation
medication. In addition, the duration, as well as the turnaround
time between procedures, affects the efficiency of
the catheterization laboratory. Despite these issues, many
congenital intervention programs continue to use the combination
of TEE and fluoroscopy as their imaging modalities
to guide ASD and PFO device closure procedures. It
remains to be determined whether the trend will be away
from TEE as the ultrasound imaging component when the
cost of newer technology and the comfort of the invasive
cardiologist interpreting the echocardiographic images
improves.
2D Transesophageal Echocardiography
Without Fluoroscopy
Transcatheter closure of ASD and PFO under TEE guidance
without fluoroscopy has been previously described. In
2000, Ewert and colleagues reported 22 patients that
underwent attempted percutaneous atrial septal device
closure with TEE alone.7 They were able to perform the
intervention in 19 patients without fluoroscopy and in
three patients in whom fluoroscopy was required.
Although they demonstrated that the procedure could be
safely performed under TEE guidance without fluoroscopy,
the patients required a general anesthetic and significantly
higher doses of sedation secondary to a longer duration of
TEE. Consequently, this imaging option exists for select
patients but is not considered by the majority of congenital invasive cardiologists as the imaging option of choice for
this patient population.
2D Intracardiac Echocardiography With Fluoroscopy
Intracardiac echocardiography (ICE) uses an echocardiogram
with an intracardiac ultrasound catheter that
can be maneuvered in the cardiac chambers in a manner
analogous to a TEE probe in the esophagus. The ICE
catheter provides a 2D pie-shaped image oriented perpendicular
to the long axis of the catheter, rather than
images of only the surrounding vessel wall, as seen with
intravascular ultrasound. The first ICE catheter used during
electrophysiology procedures was a 9-F, 9-MHz, single-
plane probe with a mechanically rotating single crystal
and radial cross-sectional imaging at 10° oblique.
Unfortunately, the radial imaging plane was a cross-sectional
image with limited tissue penetration to only 4 cm.
The next-generation catheter was a 10-F, 7.5-MHz, singleplane
phased-array linear-arc probe. This system
improved on some of the limitations in penetration, but
a detailed image of all of the cardiac anatomy remained
inadequate. Subsequently, the Acuson AcuNav 10-F (and
later 8-F), multifrequency, single-plane diagnostic ultrasound
catheters (Siemens Healthcare, Malvern, PA) were
developed with a 90-cm insertion length with tissue penetration
to approximately 12 cm. This probe is maneuvered
within the cardiovascular system using a four-way
tip deflection control knob. It provides a 2D image and
also has pulsed wave, continuous wave, color-flow mapping,
and tissue Doppler capabilities. In addition, this ICE
probe has multiple frequency capabilities ranging from
5.5 to 10 MHz. Therefore, invasive cardiologists can select
a lower frequency to allow more tissue penetration with less image resolution, or a higher frequency to give less
tissue penetration but with improved image resolution.
The 5.5-MHz frequency often allows enough tissue penetration
from the right atrium to visualize the left ventricle,
left atrial appendage, or pulmonary veins. The 10-
MHz frequency remains limited in tissue penetration but
significantly improves the image detail of the atrial septum
from the right atrium.
Procedures utilizing ICE are mainly performed with the patient under moderate sedation, and a standard 9- or 11-F sheath is placed in the femoral vein. Although placement of the two venous sheaths can be in the same or opposite femoral vein, it has been our practice (for the past 6 years) to place both sheaths in the right femoral vein (11 F x 2), with no adverse events. Subsequently, the ICE probe is advanced through the sheath (with the operators' fingers close to the tip to avoid damaging the transducer) and then carefully, under fluoroscopic guidance, to the right atrium. Slight deflection of the probe tip may be necessary to enable smooth advancement of the probe along the iliofemoral junction (rightward deflection) and past the hepatic veins (leftward deflection). The orientation of the image plane is controlled through variable deflection of the catheter tip by the control knobs. The catheter tip flexes approximately 160° in the direction of deflection. To visualize the atrial septum or the compliant sizing balloon, the probe is advanced to the high right atrium and deflected to the lateral right atrial wall (large control knob slightly toward the operator) (Figure 3). The probe is then locked in position (turn locking mechanism away from the operator), and fine movements of the probe are performed with the middle smaller control knob away from the operator to move toward the superior vena cava, and toward the operator to move toward the aorta. Continual manipulation is performed in a superior and inferior direction, or clockwise and counterclockwise rotation, resulting in a change in the image plane to enable all of the atrial septum to be visualized. If the image remains suboptimal, either the “tricuspid valve view” or the “long axis image of the ascending aorta view” can often be used as a “home view” to start the rotation.8
Although ICE can be performed safely with careful manipulation, the reported procedural adverse event rate is 1% to 3%.9,10 Adverse events include arrhythmia, allergic reaction from latex exposure to the sterile sleeve used to cover the handle and connecting cable, groin hematoma, and pericardial effusion. In addition, there are reports of retroperitoneal bleeding and venous perforation. Therefore, safety may be improved by advancing the probe under fluoroscopic guidance with the two articulation knobs in the neutral position and the lock released. Also, there is the option to use a longer venous sheath for the ICE probe, which eliminates movement through the iliofemoral system.
Despite these potential adverse events, the important advantages include the probe's proximity to intracardiac structures (especially the inferior atrial septum), the absence of any air or tissue interference with the catheter in the right atrium, and its maneuverability to achieve almost any orthogonal view. These strengths make it an ideal adjunct imaging tool for the invasive cardiologist in the congenital catheterization laboratory because it permits more accurate assessment of the congenital heart defect and real-time guidance of intervention in combination with fluoroscopy. Limitations remain, however, including the absence of multiplane views when compared to TEE, the inability to image the left atrial disc as well as TEE, 3D capabilities are not in real time, the current probe cost, and the probe size limiting manipulation as well as resulting in the need of an additional 9- to 11-F sheath. Despite these limitations, it is generally agreed that ICE is on the cutting edge of an evolving paradigm in which multiple imaging tools can improve the safely and overall outcome of atrial septal device closure.11
2D Intracardiac Echocardiography
Without Fluoroscopy
ICE has been used in select patients, such as during
pregnancy, to perform percutaneous atrial septal closure
without fluoroscopy. Schrale and colleagues reported on
three patients who experienced a neurological event during
or immediately preceding pregnancy.12 Due to evidence
of recurrent events or relative contraindication to
anticoagulation, they proceeded to percutaneous device closure during the second trimester. Successful closure
was achieved with the Helex septal occluder device in all
three patients. There were no procedural adverse events,
and all three pregnancies were successful, with the
patients remaining free of further neurologic events.
FUTURE DIRECTIONS
Real-Time Magnetic Resonance Imaging With
or Without Fluoroscopy
Although a combination of fluoroscopy and echocardiography
is the standard method for guidance during
percutaneous atrial septal device closure, there are disadvantages,
including poor soft tissue visualization and
exposure to radiation. Razavi and colleagues describe
multiple diagnostic and two interventional cardiac
catheterization procedures guided by magnetic resonance
imaging (MRI) with some fluoroscopy support.13
They combined MRI and radiographic imaging facilities
and showed that interventional cardiac catheterization
guided by MRI allowed better soft tissue visualization,
provided more pertinent physiological information, and
resulted in lower radiation exposure than fluoroscopically
guided procedures. In addition, Buecker and colleagues
tested the feasibility of applying MRI guidance
for percutaneous closure of PFO in seven piglets without
fluoroscopic assistance.14 They used a specially designed
prototype of a nonmagnetic atrial septal closure device
introduced via the femoral vein, and deployment of the
device was depicted by real-time MRI. They also reported
that the initial misplacement of the device in two
cases was easily detected and corrected. Therefore, in
this animal model, they demonstrated that real-time
MRI guidance of atrial septal closure without the use of
fluoroscopy is feasible. Consequently, MRI guidance may
evolve into a viable method for diagnostic and interventional
cardiac catheterization in patients with congenital
heart disease.
Real-Time 3D Transesophageal Echocardiography
With Fluoroscopy
Balzer and colleagues recently published their experience
with the use of real-time 3D TEE for guiding percutaneous
atrial septal device closure.15 Their series included 25
patients undergoing device placement guided by fluoroscopy
and real-time 3D TEE. The addition of real-time 3D
TEE allowed safe device deployment with no adverse
events in all patients, as well as a significant reduction in
fluoroscopy time.
CONCLUSION
Ultrasound imaging technology has advanced significantly
during the last few decades. The application of current ultrasound imaging modalities has augmented
the capabilities of invasive cardiologists performing percutaneous
ASD and PFO closure procedures. The invasive
cardiologist, rather than an echocardiographer, now
more often performs the imaging, and so this physician
must be appropriately skilled in the manipulation of the
imaging catheter as well as the interpretation of the
images generated.
ICE is clearly emerging as an important aid to the invasive cardiologist for septal closure device implantation and guidance during interventional procedures. In the majority of percutaneous atrial septal interventions, ICE eliminates the need for TEE and general endotracheal anesthesia. Despite the current limitations of cost and probe size, the eventual development of smaller catheters and multiplane, as well as real-time 3D imaging capabilities, may facilitate the use of ICE for imaging during percutaneous ASD and PFO device closure.
John F. Rhodes, Jr, MD, is Chief of Clinical Cardiology, Medical Director, Pediatric and Adult Congenital Cardiac Catheterization Lab, Duke Heart Center, Duke University Medical Center in Durham, North Carolina. He has disclosed that he is the principal investigator for the Gore REDUCE Study and is a paid consultant to W. L. Gore & Associates. Dr. Rhodes may be reached at (919) 681-2880; john.rhodes@duke.edu.
Advertisement
Advertisement