Advertisement
Advertisement
September/October 2010
Right Heart Percutaneous Valve Therapies
An update on percutaneous valve options for treating right ventricular outflow tract dysfunction in patients with congenital heart disease.
By Julie A. Vincent, MD, FACC, FSCAI, and William E. Hellenbrand, MD, FSCAI
The use of a surgically placed, valved conduit between the subpulmonary ventricle and the pulmonary artery in patients with congenital heart defects was first reported in the mid 1960s.1 Such conduits made possible the repair of many complex congenital cardiac lesions involving atresia or hypoplasia of the right ventricular outflow tract (RVOT). Patients with multiple types of congenital cardiac lesions benefited from reparative surgery involving RVOT conduits and, as technical advances in surgery occurred, these procedures were performed in younger (and smaller) patients.2-11 Although initial results were exciting, it soon became apparent that conduit failure in the form of progressive valve/conduit stenosis from calcification or intimal proliferation, and/or valvular degeneration causing progressive pulmonary regurgitation (PR), was inevitable for all types of RVOT conduits.12-21 In the growing child, somatic outgrowth of the conduit(s) occurs. Irrespective of the cause of conduit failure, the need for multiple surgical procedures for conduit revisions was to become the rule for these patients. Furthermore, some found that second and subsequent conduits had shorter freedom from failure than the originally placed conduits.15
Transcatheter therapies, such as balloon valvuloplasty/ angioplasty and/or bare-metal stenting, have been instituted to provide palliative relief for conduit stenosis in hopes of extending the functional life span of conduits and have been shown to be somewhat beneficial in prolonging interval(s) between conduit replacement( s).22-25 However, especially in the case of conduit stenting, patients are left with significant PR. Multiple long-term follow-up studies have suggested that chronic volume overload of the RV can lead to impaired systolic and diastolic function of the RV, dilation of the tricuspid valve annulus leading to significant tricuspid regurgitation, impaired exercise capacity, atrial and ventricular arrhythmias, heart failure, and an increased risk of sudden death.25-30 Furthermore, restoration of pulmonary valve competence at an appropriate time has resulted in improvement in RV function, exercise tolerance, and incidence of arrhythmias.28,30,31 Until recently, there has been no transcatheter or less-invasive treatment for PR and therefore, these patients have either been managed medically or referred back for repeat surgery.
For nearly 20 years, investigators have attempted to come up with a nonsurgical treatment for failed conduits and valves.32-34 Dr. Philipp Bonhoeffer was the first (in 2000) to perform percutaneous replacement of the pulmonary valve in a right ventricular–pulmonary artery (RV-PA) prosthetic conduit of a human with valve dysfunction. 35 Since then, rapid advancements in the development of percutaneous catheter-based therapies for cardiac valve repair have occurred.36-41
Currently, in the United States, there are only two transcatheter heart valves that have been used in the pulmonary position for RVOT dysfunction: the Medtronic Melody transcatheter pulmonary valve (Medtronic, Inc., Minneapolis, MN) and the Edwards Sapien transcatheter heart valve (THV) (Edwards Lifesciences, Irvine, CA).42-46 The basic concept of both devices involves intravascular stent and bioprosthetic valve technologies.
This article provides an update on percutaneous valves used in the treatment of RVOT conduit dysfunction. It also reviews recommended criteria for use, special procedural considerations for transcatheter valve placement in the RVOT position, and recently published outcomes regarding the use of these valves for RVOT dysfunction.
MEDTRONIC MELODY TRANSCATHETER
PULMONARY VALVE
The current Melody valve is a modification of the initial
valve used by Dr. Bonhoeffer and consists of a platinum-
iridium stent with a valved segment of natural
bovine jugular vein.34,39 The Melody valve has an initial
stent length of 28 mm and an initial diameter of 18 mm
(Figure 1A). The venous segment is sutured to the stent
around its entire circumference at inflow and outflow
and at each stent node using 5-0 polypropylene sutures.
The suture is blue at the outflow to signify the outflow
end of the valve for appropriate alignment of the valve
onto the delivery system and at implant (Figure 1B). The
valve can be crimped down to 6 mm for loading and reexpanded
up to 22 mm in diameter, depending on which
size delivery system is used.
Delivery System
The Melody valve is implanted using the Ensemble
catheter delivery system (Medtronic, Inc). The Ensemble
system is composed of a balloon-in-balloon deployment
catheter, which comes in three different sizes (18, 20, and
22 mm) based on the outer balloon diameter of the balloon-
in-balloon catheter onto which the Melody valve is
front-loaded and hand crimped. There is an outer
retractable sheath that covers the valve during delivery
and is pulled back just before deployment. The sheath
has a side arm for flushing or delivering contrast. The
entire system has a 22-F crossing profile no matter which
outer balloon diameter is used (Figure 2).
Melody Outcomes and Follow-Up Data
Initial outcomes data involving use of the Melody
valve in the pulmonary position in human subjects with
RVOT dysfunction are from Dr. Bonhoeffer's group in
Europe.35,36,39,47 In September 2006, the first prospective,
nonrandomized, multicenter trial involving the Melody
valve was initiated in the United States under an investigational
device exemption using a standardized protocol.
The objectives of the trial were to determine safety, procedural
success, and short-term effectiveness of the Melody
valve by assessing 6-month outcomes. Inclusion criteria
included patients aged more than 5 years, weight > 30 kg,
presence of a surgically placed conduit with an initial
implant diameter of > 16 mm, and evidence of conduit dysfunction
defined as either: 1) New York Heart Association
(NYHA) class II, III, IV with RVOT mean Doppler gradient by
echo > 35 mm Hg or moderate-to-severe PR; or 2) NYHA
class I with RVOT mean gradient > 40 mm Hg or severe PR
with RV dilation and dysfunction (RVFS < 40%) by echocardiography. The initial short-term results involving
the first 34 patients enrolled in the study were reported
by Zahn et al.42 The short- and medium-term outcomes
of 136 patients enrolled through August 2009 were
recently published by McElhinney et al.43 In the updated
report of these patients by McElhinney et al, the procedural
success rate was high at 98%, with only 6% experiencing
serious procedural adverse events, which included
uncontained conduit rupture in two patients, guidewireinduced
pulmonary artery perforation in two patients,
and femoral vein thrombosis in one patient. There was
one death after implantation that was not device related. This patient died 20 days after implantation after experiencing
complications due to coronary dissection with
diagnostic coronary angiography for coronary assessment
before valve implantation. Although only 15% of the
patients were categorized as NYHA class I before implantation,
at every follow-up interval after implantation at
least 75% were categorized as NYHA class I, with no
patients considered class III or IV. Before implantation,
more than 80% of the patients had moderate to severe
PR; however, at each follow-up interval out to 2 years,
more than 90% of the patients had no/trivial PR and no
patients had more than mild PR (Figure 3). Paired preimplantation
and 6-month postimplantation cardiac MRI
data were consistent with a significant decrease in RV
end-diastolic volumes, RV mass, and PR fractions at follow-
up. Stent fractures were observed in 26% of the
patients, with the majority occurring in patients who did
not undergo prestenting of their conduit before valve
implantation. Most of these patients had been enrolled
early on in the study when concomitant procedures (such
as prestenting of the conduit) were not allowed. McElhinney
et al did report that 10 of 122 patients (8%) required reintervention
during follow-up, with all reinterventions performed
due to increased RVOT gradients. All but one of
these patients were noted to have stent fractures at the
time of reintervention. Initial reintervention was balloon
angioplasty in two patients and placement of a second
Melody valve in eight patients. Only one patient required
surgical conduit replacement at follow-up, and late mortality
was zero.
EDWARDS SAPIEN TRANSCATHETER
HEART VALVE
The Edwards Sapien (formerly known as the Cribier-
Edwards) THV was initially developed as a percutaneous
valve option for aortic valve disease and was first successfully
placed in a human for calcified aortic stenosis by
Cribier.37 In 2005, a previous version of the Sapien valve
was used under a compassionate use protocol for percutaneous
placement in an obstructed RV-PA homograft of
a patient who had previously undergone a Ross procedure.
44 Most recently, the Edwards Sapien THV was used
in the treatment of seven patients with RV-PA conduit
failure.45 The Sapien valve also combines balloon-expandable
stent and bioprosthetic valve technology. The frame
is a radiopaque, stainless steel, expandable support structure
with an integrated, unidirectional, trileaflet, tissue
valve fabricated from three equal sections of bovine pericardium.
The valve leaflets are hand-sewn to the stent
frame (Figure 4). The valve is available in 23-mm- and
26-mm-diameter sizes, with heights of 14.5 and 16 mm,
respectively.45
Delivery System
The Sapien valve has been delivered via a transfemoral
route using the RetroFlex 3 delivery system (Edwards
Lifesciences). The RetroFlex 3 system consists of the introducer
sheath set, a dilator kit, and the RetroFlex balloon
catheters. The delivery system is an articulating “flex”
catheter with a handle that provides a rotational grip for
articulation of the distal portion of the catheter, a
tapered tip at the distal end of the delivery system to
facilitate tracking of the Sapien valve into the RVOT, and
a balloon for deployment of the valve. This valve requires
a specialized crimper that symmetrically reduces the
overall diameter of the valve from its expanded size to its
collapsed (mounting) size on the delivery balloon catheter.
PROCEDURAL CONSIDERATIONS
Coronary Artery Testing
A few comments should be made regarding the risk of
coronary artery compression with stent and/or valve
placement in the RVOT position, because this is the most
common reason why patients who would otherwise
meet criteria for a percutaneous valve do not receive one.
As has been discussed in previous studies, it is imperative
that assessment for possible coronary compression be
performed before placing a stent and/or valve in the
RVOT conduit.24,42,48,49 In the patient population requiring
RVOT conduits, the respective cardiac defects may
either have anomalous coronary artery anatomy, such as
a single coronary giving rise to significant branches that
cross over the front of the heart, and/or the aorta may
be in such an anterior or rotated position that even normal
origins of the coronaries may still be relatively displaced.
Still in other patients, surgery for their congenital
cardiac defects may include reimplantation of the coronary
arteries, such as in an arterial switch operation or a
Ross procedure. The reimplanted coronaries, particularly
the left coronary artery, may be at risk of compression
during RVOT conduit stenting or valve placement,
depending on the postsurgical anatomic relationship
between the more anteriorly placed RV-PA conduit and
the coronaries. Often, it is not possible to predict the risk
of coronary compression with noninvasive testing; therefore,
the risk of coronary compression must be evaluated
at catheterization before bare-metal stenting and/or
valve placement. Initial screening is done by aortic root
or selective coronary angiography with a wire or catheter
in the RV-PA conduit. If there is any concern of a close
anatomic relationship between the conduit and a significant
coronary branch, simultaneous inflation of a balloon
within the conduit and aortic root and/or selective coronary
angiography should be performed (Figure 5). A few
important technical considerations include: (1) be sure the
appropriate balloon size and inflation pressure is used to
best approximate the resultant diameter of the conduit
after stent/valve implantation; (2) multiple imaging projections
may be necessary to provide the best assessment of
the conduit-coronary relationship; (3) remember that
selective coronary angiography may mask ostial stenosis or
very proximal obstruction during simultaneous conduit
balloon dilation; (4) use very dilute contrast to inflate the
balloon because a more concentrated mixture may make
it difficult to see the course of the coronary arteries behind
or underneath the balloon; (5) make sure the widest part
of the balloon being used to dilate the conduit is actually
positioned over the region or coronary in question; and
(6) remember that ringed conduits or bioprosthetic valve
rings do not remove the risk of coronary compression.
Conduit Rupture
It is not uncommon for contained conduit tears to
occur with balloon dilation for conduit stenosis. Most
often these tears are hemodynamically insignificant. In
the United States investigational device exemption trial involving the Melody valve, uncontained conduit ruptures
occurred in two patients. In the first instance, the
patient was temporarily palliated with a chest tube and
then underwent emergent surgical conduit replacement.
The second patient was treated under emergent use protocol
with a large-diameter, covered stent. Of note, the
Melody valve itself functions as a covered stent, with the
venous wall segment attached to the stent frame; however,
ruptures of the conduit may not be in the same region
that optimal valve placement would be and/or the ruptures
may be longer than the postimplantation valve
length.
Pulmonary Artery Injury
Pulmonary artery injury (perforation) is another serious
procedural adverse event noted in the Melody trial.
Vessel injury was presumed to have occurred due to wire
perforation. Percutaneous valve implants require use of
very stiff guidewires and great care should be taken in
attaining appropriate and stable wire position before
valve deployment to keep the risk of vessel injury to a
minimum.
PREPARATION OF CONDUITS
FOR VALVE IMPLANT
Despite improved stent technologies over the years,
stent fractures are a common complication after implanting
intravascular stents into cardiovascular-related positions.
23,24,50-53 This has been shown to be true for the
Melody valve, as well.43,47,54,55 The etiology of stent fractures
is thought to be secondary to the dynamic implant
environment and the intrinsic metal characteristics of the
stents. With ongoing implantation experience, the Melody
investigators have implemented certain practices to prepare
conduits for implantation in the hopes of decreasing
the incidence of stent fractures. These include: (1) aggressive
predilation of calcified and stenotic conduits using
high-pressure balloons and a balloon diameter < 110% of
the conduit diameter at implantation; (2) prestenting conduits
with bare-metal stent(s) until evidence of recoil of
the conduit during balloon deflation is gone; and (3)
attempting to implant the valve in a position that is not
directly under the sternum, when possible. Preliminary
evaluation of the Melody data suggests that when these
techniques are implemented, the incidence of stent fractures
at follow-up is significantly decreased, with 35% of
fractures occurring in conduits not prestented compared
to only 7% of fractures occurring in prestented conduits.
When valve stent fracture occurs and reintervention is
indicated, placement of bare-metal stent(s) within the initially
placed valve with subsequent placement of a second
valve has been shown to be successful.43,47
DEVICE AVAILABILITY
IN THE UNITED STATES
On January 25, 2010, the United States Food and
Drug Administration formally approved the Melody
valve for placement in dysfunctional RVOT conduits
under a Humanitarian Device Exemption, making it the
first transcatheter heart valve to receive United States
Food and Drug Administration approval. Enrollment
into the United States feasibility trial involving use of
the Edwards Sapien THV in the pulmonic position was
completed in May 2010. Plans for a pivotal study are in
progress.
CONCLUSION
Percutaneous valve therapies for RVOT conduit dysfunction
have been shown to be feasible, safe, and effective,
with high procedural success rates and low adverse
event rates. Procedural considerations, such as careful
assessment for risk of coronary compression, cautious
use of stiff guidewires, and awareness of the potential
for serious adverse events such as conduit rupture, as
well as knowing how to prevent and manage such
events, will lead to safe and successful percutaneous
valve implantations.
Julie A. Vincent, MD, FACC, FSCAI, is Director of the Cardiac Catheterization Laboratories and Director of the Pediatric Cardiology Fellowship Program at Morgan Stanley Children's Hospital of New York-Presbyterian and the Komansky Center for Children's Health at Weill Cornell Medical Center. She is Assistant Professor of Clinical Pediatrics at Columbia University, College of Physicians & Surgeons in New York, New York. She has disclosed that she is a consultant to and proctor for Medtronic, Inc. Dr. Vincent may be reached at (212) 305-6069; jav2136@columbia.edu.
William E. Hellenbrand, MD, FSCAI, is the Director of the Division of Pediatric Cardiology, a combined program at Morgan Stanley Children's Hospital of New York- Presbyterian and the Komansky Center for Children's Health at Weill Cornell Medical Center. He is Professor of Clinical Pediatrics at Columbia University, College of Physicians & Surgeons in New York, New York; and holds the position of the Welton M. Gersony Professor of Clinical Pediatric Cardiology. He has disclosed that he is a consultant to and proctor for Medtronic, Inc. Dr. Hellenbrand may be reached at (212) 342-0610; wh148@columbia.edu.
Advertisement
Advertisement