The Next Wave of Mechanical Circulatory Support Devices

Next-generation mechanical circulatory support devices in different phases of development and clinical use are anticipated to improve outcomes and quality of life in patients with advanced heart failure.

By Carlo R. Bartoli, MD, PhD; and Robert D. Dowling, MD, PhD (Hon)
 

The recent success of continuous-flow (nonpulsatile) circulatory support devices has markedly increased long-term artificial circulation as a treatment option for patients with advanced heart failure. Multiple continuous-flow devices are clinically approved as a bridge to transplant and destination therapy. However, morbidities related to device implantation via median sternotomy with cardiopulmonary bypass and frequent adverse events, such as acquired von Willebrand syndrome, gastrointestinal bleeding, device thrombosis, progressive aortic insufficiency, and stroke, complicate management and impose further morbidity and mortality. As a result, therapy with these devices is still limited to only the sickest patients. The number of patients referred for durable circulatory support device implantation is small compared to the number of patients with advanced heart failure who could benefit from prolonged circulatory support with fewer adverse events. Therefore, there is an urgent need for next-generation devices with less-invasive implantation approaches, novel mechanisms of blood flow, physiologic performance, improved hemocompatibility, and fewer adverse events. This article highlights next-generation mechanical circulatory support (MCS) devices in different phases of development and clinical use that are anticipated to improve outcomes and quality of life in patients with advanced heart failure.

background

Survival with advanced heart failure remains poor. Cardiac transplantation offers the best opportunity for long-term survival. However, limited availability of donor hearts, strict selection criteria, insurance constraints, and complications of lifelong immunosuppression restrict heart transplantation to only 7,000 patients annually worldwide. As a result, over the past 20 years, MCS has emerged as a standard long-term therapy for adult patients with intractable heart failure.1,2

Over the past decade, first-generation pulsatile devices, which mimic native cardiac function, were replaced by rotary blood pumps that continuously unload the failing heart.3 So-called continuous-flow left ventricular assist devices (LVADs) have markedly improved patient outcomes.1,2 Continuous-flow devices are smaller, more reliable and durable, more energy efficient, and less traumatic to implant than first-generation devices. However, continuous-flow devices have introduced a new nonpulsatile physiology with unforeseen complications. Blood trauma from supraphysiologic shear stress in these devices causes acquired von Willebrand syndrome in all continuous-flow LVAD patients,4 gastrointestinal angiodysplasia and bleeding in 20% to 40% of patients,4-6 and device thrombosis in 2% to 13% of patients.7,8 Nonpulsatile flow also contributes to progressive de novo aortic valve insufficiency in 37%9 and stroke in close to 15% of patients.10 To reduce adverse events and improve patient outcomes, refinement of current pump designs11 and development of novel devices12 are needed.

The following sections discuss next-generation LVADs and total artificial hearts (TAHs) that are in various phases of development and clinical use. We hope that increased awareness of the relationships that exist between circulatory support device implantation strategy, design features, operational and flow characteristics, and adverse events stimulate enthusiasm in the community for further development and availability of less-invasive, safer devices.

NEXT-GENERATION CONTINUOUS-FLOW LVADs DESIGNED TO REDUCE ADVERSE EVENTS

Continuous-flow LVADs have achieved overall survival of 80% at 1 year and 70% at 2 years.13 However, the current profile and frequency of adverse events restrict these devices to only the sickest patients. Better clinical outcomes are necessary before prolonged LVAD support is more widely accepted.

Continuous-flow devices contain an impeller that rotates at 1,500 to 30,000 rpm to generate forward flow. Blood courses through narrow gaps (50–500 µm) at high velocity. As a result, shear stress may exceed physiologic values by one to two orders of magnitude.1,2 For example, the industry standard, HeartMate II (Abbott Vascular), generates peak shear stress > 1,500 Pa14 (normal physiologic intravascular shear stress is approximately 2–8 Pa15). Supraphysiologic shear stress causes blood trauma. The foremost types of LVAD-associated blood trauma with known clinical consequences include von Willebrand factor degradation3 (which contributes to bleeding), platelet activation4 (which contributes to thrombosis), and subclinical hemolysis5 (which contributes to thrombosis). As a result, LVAD hemocompatibility, a term used to characterize the clinical impact of biophysical interactions and blood trauma at the device-blood interface, is gaining attention as a major area for improvement of future-generation LVADs.

Also problematic, current-generation continuous-flow LVADs operate at a fixed rpm without physiologic control. Devices do not sense physiologic feedback to adjust pressure and flow according to preload, afterload, intracardiac hemodynamics, or metabolic demands. Other major areas of focus for next-generation LVADs are design features and physiologic control algorithms to promote load sensitivity, optimize unloading, and facilitate device weaning.

Currently approved LVADs and TAHs contain polymeric valves and diaphragms or mechanical contact bearings that wear down. To counter this limitation and increase device durability, third-generation devices include a hydrodynamically or magnetically suspended bearingless impeller that does not wear over time or generate frictional heat.16

Lack of pulsatile blood flow has also raised concerns. Consequently, continuous-flow pulsation algorithms are being developed to generate pulsatility.17,18 As device companies address these issues associated with new systems, continuous-flow LVAD therapy is likely to further increase in clinical success and prevalence.

Figure 1. The Evaheart 2 device.

Evaheart 2

Evaheart, Inc. has developed a novel centrifugal-flow LVAD (Figure 1). The original Evaheart device has been available in Japan since 2010. Outcomes in the initial clinical trial were favorable.19 Recently, the next-generation Evaheart 2 device was implanted in a patient in Japan, and it is poised for clinical trial in the United States.

Evaheart devices have an open-vane hydrodynamic impeller design with large blood flow gaps. The result is preserved aortic pulsatility, low operational rpm, low shear stress, minimal blood trauma, and low likelihood of thrombosis. Ex vivo experiments demonstrated significantly less blood trauma than with currently approved continuous-flow LVADs.11 As a result, patients with the Evaheart device experience a remarkably low incidence of gastrointestinal bleeding.19 This is a clear demonstration of the impact of device design on clinical outcomes.

Evaheart 2 is a smaller version of the first-generation device with the same open-vane centrifugal-flow design and a novel inflow cannula. The new inflow cannula was redesigned with a double cuff to minimize protrusion into the left ventricle, eliminate local blood stagnation, and prevent inflow cannula malposition. This inflow cannula may provide a complete solution to inflow thrombus formation and ventricular wall suction, which are root causes of ischemic stroke and pump thrombosis. Further clinical experience with this device will define potential advantages of preserved pulse pressure, minimal blood trauma, and improved cannula design.

Figure 2. The HeartMate 3 device.

HeartMate 3

The HeartMate 3 device (Abbott Vascular) is a magnetically levitated centrifugal-flow LVAD (Figure 2). A recent multicenter, randomized clinical trial, MOMENTUM 3, compared the HeartMate 3 to its predecessor, the HeartMate II. At 6 months20 and at 2 years,2 rates of death or disabling stroke were the same with each device. However, there were significantly fewer pump malfunctions and no device thrombosis with the HeartMate 3 device.

HeartMate 3 is implanted via median sternotomy or a less-invasive sternal-sparing approach21 on cardiopulmonary bypass. Favorable design features include a magnetically suspended impeller that is intended to provide continuous support for a decade with low power consumption. An artificial pulse mode may have numerous beneficial effects that include improved myocardial reverse remodeling, prevention of aortic valve thrombosis, leaflet fusion, de novo aortic insufficiency, reduced arterial stiffening, better end-organ function, and reduced bleeding.17,22

LESS-INVASIVE CONTINUOUS-FLOW LVADs

If combined, the benefits of a minimally invasive surgical approach and durable flow augmentation may expand the potential patient population for LVAD therapy. As LVADs are miniaturized, minimally invasive implantation may increase acceptance by physicians, who are more likely to refer patients for less-invasive surgical therapies.23 As a result, earlier intervention in less-sick patients may increase the public health impact of MCS. Novel surgical approaches include limited thoracotomy, subxiphoid access, placement of support devices in the infraclavicular fossa, or percutaneous implantation.

Furthermore, certain minimally invasive and percutaneous operative approaches do not require cardiopulmonary bypass.24 As a result, less coagulopathy may reduce postoperative bleeding and blood transfusions, which play a role in right ventricular failure and infection with LVADs.25

Figure 3. The Impella 5.5 device.

Impella 5.5

The Impella family of devices (Abiomed, Inc.) have been developed and approved by the FDA to provide acute circulatory support for high-risk percutaneous coronary intervention (PCI) and cardiogenic shock (Figure 3). These catheter-based devices are percutaneously placed retrograde across the aortic valve with the pump inflow in the left ventricle and the pump outflow in the proximal ascending aorta.

Recently, the Impella 5.5 device underwent initial human implantation in Europe (n = 4). This device is implanted through the subclavian artery and provides up to 6 L/min of flow for up to 30 days. Early ambulation and discharge to home are possible. Optical sensors in the device display real-time aortic and left ventricular pressure waveforms to confirm appropriate device placement and optimize hemodynamics and device weaning. Design modifications to extend the duration of support to > 1 year are in advanced developmental stages.

The potential advantages of a full-support device implanted without a sternotomy or cardiopulmonary bypass and provides real-time intracardiac hemodynamic data to guide weaning and myocardial recovery are significant.

Figure 4. The Aortix.

Aortix

Aortix (Procyrion, Inc.) is a catheter-deployed, intra-aortic, continuous-flow device for the management of acute decompensated heart failure with cardiorenal syndrome. It features a miniature rotary pump that is caged within a catheter-based nitinol strut system (Figure 4). The device is percutaneously deployed in the descending aorta and unloads the left ventricle by decreasing proximal aortic resistance while providing distal aortic flow augmentation.

Preclinical studies demonstrated effective deployment and retrieval. In an ovine heart failure model, the Aortix device decreased cardiac energy consumption and improved cardiac and urine output.12 Initial patients supported during high-risk PCI (n = 6) also demonstrated significant increases in cardiac and urine output (J. Heuring, PhD, written communication, December 2018). Further clinical studies will define the role of the Aortix device in acute and chronic heart failure.

Figure 5. The iVAS device.

iVAS

The iVAS intraventricular assist system (NuPulseCV, Inc.) is a novel, counterpulsation device similar to an intra-aortic balloon pump (Figure 5). An FDA-approved feasibility study was performed in patients with advanced systolic heart failure. All participants (n = 13) were successfully bridged to cardiac transplantation with no deaths or neurologic events.26 Improved biventricular function was observed.27

The iVAS balloon is implanted in the aorta via left axillary artery cutdown. The 50-mL balloon inflates during diastole to augment coronary blood flow and deflates during systole to reduce afterload. A limited operative approach combined with the proven benefits of counterpulsation therapy is desirable and may have significant clinical impact for ambulatory support in patients with less-advanced heart failure.

NEXT-GENERATION CIRCULATORY SUPPORT DEVICES WITH NOVEL FLOW MECHANISMS and better hemocompatibility

Strong evidence suggests that limited hemocompatibility with current-generation LVADs contributes to LVAD-associated bleeding, thrombosis, and stroke.5-7,9 These complications cause significant morbidity and mortality and have limited the role of LVADs in patients with less-advanced stages of heart failure.10 As such, it is becoming increasingly clear that next-generation LVADs with less blood trauma (better hemocompatibility) are urgently needed to reduce adverse events.

Additionally, current-generation continuous-flow LVADs operate at a fixed rpm. Devices are unable to automatically respond to changing hemodynamics and metabolic demands. Toward this goal, next-generation devices described in this section have been specifically designed to improve hemocompatibility and physiologic responsiveness.

Figure 6. The TORVAD device.

TORVAD

The TORVAD toroidal ventricular assist device (Windmill Cardiovascular Systems, Inc.) is a novel, positive-displacement, toroidal-flow LVAD (Figure 6).28-30 The toroidal-flow mechanism creates pulsatile blood flow and allows physiologic device control with low shear stress28 and minimal blood trauma.31 At 40 to 260 rpm, the 30-mL stroke volume generates 1 to 8 L/min of flow. Synchronous counterpulsation or asynchronous modes may be used to optimize ventricular unloading and facilitate device weaning in the setting of myocardial recovery.

The pumping mechanism of TORVAD is distinctly different from first-generation, sac-type, pulsatile LVADs. To simultaneously fill and eject, TORVAD spins one of two magnetic pistons within a torus (doughnut-shaped) chamber to generate unidirectional pulsatile flow. While one piston is temporarily fixed as a “virtual valve,” the other spins around the torus to generate flow through a 1.7-cm-wide path comparable to a large artery. At 5 L/min and 70 mm Hg, TORVAD generates shear stress of approximately 10 Pa,29 which is near physiologic values.

In ex vivo experiments with human blood, TORVAD caused no platelet activation or hemolysis and minimal von Willebrand factor degradation.31 These data are consistent with findings from 60-day bovine implants in which hemolysis and von Willebrand factor degradation were not significant.32 Animals were not anticoagulated, and thromboembolism was not observed.

Furthermore, TORVAD senses pressures within the pump and can automatically adjust pump output to optimize hemodynamic support. For these reasons, toroidal flow is a promising new LVAD design with pulsatile flow, physiologic device control, and significantly less blood trauma.

Figure 7. The Neptune device.

Neptune

Neptune (CorWave) is a novel device that employs wave membrane physics to generate pulsatile blood flow (Figure 7). The wave membrane (a flexible biomimetic polymer) mimics the undulating motion of a fish tail. The membrane is driven by electromagnets that create membrane oscillations to generate forward propulsion of blood. Physiologic pulsatility (> 35 mm Hg) and shear stress are observed.

The Neptune device is a full-support device implanted via median sternotomy that generates 5 to 6 L/min of blood flow. Acute and chronic animal studies demonstrated cardiac output of 5 L/min at physiologic blood pressures when the pump was synchronized with native heart function. After 9 days of support, no evidence of pump thrombus or thromboembolism was observed (C. Botterbusch, written communication, December 2018). Final design of the pump and external components will be followed by additional preclinical implants. First-in-human implantations are planned for 2020.

NOVEL TOTAL ARTIFICIAL HEARTS

Despite recent advancements in LVAD technology, the management of biventricular failure continues to be a major challenge. Although survival with LVADs continues to improve, there remains significant early morbidity and mortality due to right ventricular failure.33,34 In select patients, a TAH may be the best treatment option. The CardioWest (SynCardia Systems, LLC) is the only artificial heart available for clinical use in the United States. However, a major limitation is that the CardioWest does not fit in many men and most women.

The shift from volume displacement pumps to continuous-flow devices has progressively decreased size and increased the durability of LVADs. Not surprisingly, the development of continuous-flow TAHs is underway.

Figure 8. The BiVACOR TAH.

BiVACOR

BiVACOR, Inc. has developed a novel, continuous-flow TAH (Figure 8). Unlike previous-generation TAHs, the BiVACOR TAH uses rotary pump technology to provide biventricular support. BiVACOR consists of a single, magnetically levitated spinning disc positioned between separate right and left ventricular chambers. On either side of the disc are impeller blades that simultaneously generate independent flow from each ventricle. A unique feature of the device allows the disc’s relative axial position to change according to the loading conditions of each ventricle. This design allows for passive-flow adaptation with a Frank-Starling–like response and dynamic control of right-left output balance.35 For example, an increase in right ventricular preload pushes the disc to the left and increases right ventricular output and vice versa. Magnetic levitation eliminates contact bearings, and the continuous-flow mechanism precludes unidirectional valves for an expected durability of 10 years of support.

BiVACOR is orthotopically implanted after native cardiectomy. Unlike previous-generation TAHs, the smaller BiVACOR will accommodate smaller patients (body surface area > 1.2 m2). In chronic calf implantations, the BiVACOR generated peak and mean flows of 17 and 10 L/min, respectively. A pulsatile flow algorithm generated a pulse pressure of 40 mm Hg.36 Animals exhibited normal end-organ function. Additional animal studies are underway to support an application for a human clinical trial.

Biventricular replacement with physiologic hemodynamics, right-left balance, and physiologic control for prolonged circulatory support may have a significant impact for patients with life-threatening biventricular failure.

Figure 9. The RealHeart TAH.

RealHeart

RealHeart (Scandinavian Real Heart AB) is a novel, pulsatile TAH designed to mimic native heart function (Figure 9). RealHeart consists of two independently operated piston pumps. Each pump has an atrium and a ventricle separated by an atrioventricular valve. Movement of the atrioventricular plane toward the atria decreases ventricular pressure, which opens each valve and fills each ventricle. Movement of the atrioventricular plane toward the ventricles closes each atrioventricular valve and triggers ventricular ejection.

Extensive anatomic fit studies have been performed. In an acute porcine model, RealHeart delivered a wide range of cardiac outputs with a pulsatile waveform similar to the flow pattern of the native heart.37

Carmat

Carmat has developed a novel, pulsatile TAH. The Carmat device consists of two chambers in which a bovine pericardium membrane separates a blood compartment from a hydraulic fluid compartment. Electrohydraulic pressurization of the membranes ejects blood through unidirectional bioprosthetic valves. Bioprosthetic blood-contacting surfaces are anticipated to improve hemocompatibility and preclude systemic anticoagulation. Embedded sensors interact with a control algorithm that responds to changes in preload and afterload to accommodate physiologic demands.

Preclinical bovine implantations with the Carmat device demonstrated the absence of significant blood trauma with a normal profile of von Willebrand factor and normal levels of clotting factors.38 Initial clinical implantations (n = 4) allowed two patients to be discharged home.39 The clinical trial is ongoing in Europe.

CONCLUSION

Continuous-flow MCS devices have significantly improved outcomes in patients with end-stage heart failure. However, frequent device-related complications still occur. Refinement of device design is needed to improve outcomes. Miniaturization and less-invasive surgical approaches may expand the role of MCS devices. Pulsatility and physiologic device control are needed to improve device performance, recreate native physiology, promote myocardial recovery, and facilitate device weaning. Lower shear stress and less blood trauma may improve hemocompatibility and reduce adverse events. Smaller, more durable TAHs may expand the treatment options for patients with end-stage biventricular failure. Careful examination of experiences with next-generation circulatory support devices will determine the relative utility of each device.

1. Slaughter MS, Rogers JG, Milano CA, et al. Advanced heart failure treated with continuous-flow left ventricular assist device. N Engl J Med. 2009;361:2241-2251.

2. Mehra MR, Goldstein DJ, Uriel N, et al. Two-year outcomes with a magnetically levitated cardiac pump in heart failure. N Engl J Med. 2018;378:1386-1395.

3. Bartoli CR, Giridharan GA, Litwak KN, et al. Hemodynamic responses to continuous versus pulsatile mechanical unloading of the failing left ventricle. ASAIO J. 2010;56:410-416.

4. Bartoli CR, Restle DJ, Zhang DM, et al. Pathologic von Willebrand factor degradation with a left ventricular assist device occurs via two distinct mechanisms: mechanical demolition and enzymatic cleavage. J Thorac Cardiovasc Surg. 2015;149:281-289.

5. Kang J, Hennessy-Strahs S, Kwiatkowski P, et al. Continuous-flow LVAD support causes a distinct form of intestinal angiodysplasia. Circ Res. 2017;121:963-969.

6. Bartoli CR, Zhang DM, Hennessy-Strahs S, et al. Clinical and in vitro evidence that left ventricular assist device-induced von Willebrand factor degradation alters angiogenesis. Circ Heart Fail. 2018;11:e004638.

7. Bartoli CR, Zhang D, Kang J, et al. Clinical and in vitro evidence that subclinical hemolysis contributes to LVAD thrombosis. Ann Thorac Surg. 2018;105:807-814.

8. Bartoli CR, Ailawadi G, Kern JA. Diagnosis, nonsurgical management, and prevention of LVAD thrombosis. J Card Surg. 2014;29:83-94.

9. Gasparovic H, Kopjar T, Saeed D, et al. De novo aortic regurgitation after continuous-flow left ventricular assist device implantation. Ann Thorac Surg. 2017;104:704-711.

10. Purohit SN, Cornwell WK 3rd, Pal JD, et al. Living without a pulse: the vascular implications of continuous-flow left ventricular assist devices. Circ Heart Fail. 2018;11:e004670.

11. Bartoli CR, Kang J, Zhang D, et al. Left ventricular assist device design reduces von Willebrand factor degradation: a comparative study between the HeartMate II and the EVAHEART left ventricular assist system. Ann Thorac Surg. 2017;103:1239-1244.

12. Bartoli CR, Dowling RD. The future of adult cardiac assist devices: novel systems and mechanical circulatory support strategies. Cardiol Clin. 2011;29:559-582.

13. Kirklin JK, Pagani FD, Kormos RL, et al. Eighth annual INTERMACS report: special focus on framing the impact of adverse events. J Heart Lung Transplant. 2017;36:1080-1086.

14. Thamsen B, Mevert R, Lommel M, et al. A two-stage rotary blood pump design with potentially lower blood trauma: a computational study. Int J Artif Organs. 2016;39:178-183.

15. Resnick N, Yahav H, Shay-Salit A, et al. Fluid shear stress and the vascular endothelium: for better and for worse. Prog Biophys Mol Biol. 2003;81:177-199.

16. Hoshi H, Shinshi T, Takatani S. Third-generation blood pumps with mechanical noncontact magnetic bearings. Artif Organs. 2006;30:324-338.

17. Bourque K, Dague C, Farrar D, et al. In vivo assessment of a rotary left ventricular assist device-induced artificial pulse in the proximal and distal aorta. Artif Organs. 2006;30:638-642.

18. Khalil HA, Kerr DT, Schusterman MA 2nd, et al. Induced pulsation of a continuous-flow total artificial heart in a mock circulatory system. J Heart Lung Transplant. 2010;29:568-573.

19. Saito S, Yamazaki K, Nishinaka T, et al. Post-approval study of a highly pulsed, low-shear-rate, continuous-flow, left ventricular assist device, EVAHEART: a Japanese multicenter study using J-MACS. J Heart Lung Transplant. 2014;33:599-608.

20. Mehra MR, Naka Y, Uriel N, et al. A fully magnetically levitated circulatory pump for advanced heart failure. N Engl J Med. 2017;376:440-450.

21. Wood KL, Ayers B, Sagebin F, et al. Complete sternal-sparing HeartMate 3 implantation: a case series of ten consecutive patients [published online November 13, 2018]. Ann Thorac Surg.

22. Bartoli CR, Atluri P. Do patients with a continuous-flow left ventricular assist device benefit from induced-pulsatility or are we just spinning our wheels? 2015;150:945-946.

23. Casula R, Athanasiou T, Foale R. Recent advances in minimal-access cardiac surgery using robotic-enhanced surgical systems. Expert Rev Cardiovasc Ther. 2004;2:589-600.

24. Frazier OH. Implantation of the Jarvik 2000 left ventricular assist device without the use of cardiopulmonary bypass. Ann Thorac Surg. 2003;75:1028-1030.

25. Hunt SA. Mechanical circulatory support: new data, old problems. Circulation. 2007;116:461-462.

26. Jeevanandam V, Song T, Onsager D, et al. The first-in-human experience with a minimally invasive, ambulatory, counterpulsation heart assist system for advanced congestive heart failure. J Heart Lung Transplant. 2018;37:1-6.

27. Imamura T, Juricek C, Song T, et al. Improvement in biventricular cardiac function after ambulatory counterpulsation. J Card Fail. 2019;25:20-26.

28. Letsou GV, Pate TD, Gohean JR, et al. Improved left ventricular unloading and circulatory support with synchronized pulsatile left ventricular assistance compared with continuous-flow left ventricular assistance in an acute porcine left ventricular failure model. J Thorac Cardiovasc Surg. 2010;140:1181-1188.

29. Gohean JR, Larson ER, Hsi BH, et al. Scaling the low-shear pulsatile TORVAD for pediatric heart failure. ASAIO J. 2017;63:198-206.

30. Gohean JR, George MJ, Chang KW, et al. Preservation of native aortic valve flow and full hemodynamic support with the TORVAD using a computational model of the cardiovascular system. ASAIO J. 2015;61:259-265.

31. Bartoli CR, Hennessy-Strahs S, Gohean JR, et al. A novel toroidal-flow left ventricular assist device minimizes blood trauma: implications of improved ventricular assist device hemocompatibility [published online December 23, 2018]. Ann Thorac Surg.

32. Gohean JR, Larson ER, Longoria RG, et al. Low shear and thromboresistance in the synchronous pulsatile adult and pediatric TORVAD. Presented at: 63rd Annual ASAIO Conference; June 21–24, 2017; Chicago, IL.

33. Potapov EV, Loforte A, Weng Y, et al. Experience with over 1000 implanted ventricular assist devices. J Card Surg. 2008;23:185-194.

34. Kavarana MN, Pessin-Minsley MS, Urtecho J, et al. Right ventricular dysfunction and organ failure in left ventricular assist device recipients: a continuing problem. Ann Thorac Surg. 2002;73:745-750.

35. Ng BC, Smith PA, Nestler F, et al. Application of adaptive starling-like controller to total artificial heart using dual rotary blood pumps. Ann Biomed Eng. 2017;45:567-579.

36. Kleinheyer M, Timms DL, Tansley GD, et al. Rapid speed modulation of a rotary total artificial heart impeller. Artif Organs. 2016;40:824-833.

37. Szabó Z, Holm J, Najar A, et al. Scandinavian Real Heart (SRH): 11 implantations as total artificial heart (TAH)–experimental update. J Clin Exp Cardiol. 2018;9:578.

38. Smadja DM, Susen S, Rauch A, et al. The Carmat bioprosthetic total artificial heart is associated with early hemostatic recovery and no acquired von Willebrand syndrome in calves. J Cardiothorac Vasc Anesth. 2017;31:1595-1602.

39. Latrémouille C, Carpentier A, Leprince P, et al. A bioprosthetic total artificial heart for end-stage heart failure: results from a pilot study. J Heart Lung Transplant. 2018;37:33-37.

Carlo R. Bartoli, MD, PhD
Division of Cardiovascular Surgery
Hospital of the University of Pennsylvania
Philadelphia, Pennsylvania
carlo.bartoli@uphs.upenn.edu
Disclosures: Advisory board for Abiomed, Inc. and Evaheart, Inc.

Robert D. Dowling, MD, PhD (Hon)
Department of Surgery
Penn State Health
Hershey, Pennsylvania
rdowling@pennstatehealth.psu.edu
Disclosures: Advisory board for Abiomed, Inc.

 

Contact Info

For advertising rates and opportunities, contact:
Craig McChesney
484-581-1816
cmcchesney@bmctoday.com

Stephen Hoerst
484-581-1817
shoerst@bmctoday.com

Charles Philip
484-581-1873
cphillip@bmctoday.com

About Cardiac Interventions Today

Cardiac Interventions Today (ISSN 2572-5955 print and ISSN 2572-5963 online) is a publication dedicated to providing comprehensive coverage of the latest developments in technology, techniques, clinical studies, and regulatory and reimbursement issues in the field of coronary and cardiac interventions. Cardiac Interventions Today premiered in March 2007 and each edition contains a variety of topics in a flexible format, including articles covering various perspectives on current clinical topics, in-depth interviews with expert physicians, overviews of available technologies, industry news, and insights into the issues affecting today's interventional cardiology practices.