Forward-Looking IVUS in Chronic Total Occlusions

A new approach to an old problem.

By Jason H. Rogers, MD
 

Chronic total occlusions (CTOs) are frequently encountered in clinical practice and are present in approximately 30% of diagnostic coronary angiograms.1 Successful percutaneous revascularization of CTOs has been shown to result in improved left ventricular function and angina relief, decreased mortality, and freedom from subsequent coronary artery bypass grafting.2-4 However, revascularization of less than 10% of CTOs are generally attempted and represent a minority (5%–15%) of interventional procedures currently performed in the United States. The reasons why CTO procedures are rarely attempted are because they are time-intensive and can result in significant contrast and radiation exposure. In addition, complications may include vessel dissection, perforation, guide injury, embolization, myocardial infarction, and, rarely, death. Although numerous novel techniques have emerged to improve CTO crossing, many procedures still fail due to an inability to re-enter the distal true lumen. Even with advanced wire skills, maintaining or re-entering a true lumen location can be a formidable challenge. Forward-looking intravascular ultrasound (FL-IVUS) holds particular promise because it has the potential to visualize the vessel, plaque morphology, and true and false lumens in front of the imaging catheter. This enhanced visualization could be used to improve CTO crossing by continually maintaining and directing the catheter or wire toward the true lumen.

USE AND LIMITATIONS OF CONVENTIONAL IVUS IN CTOS
The current use of IVUS, specific to CTO intervention, employs two basic techniques: (1) IVUS identification and differentiation of true vessel lumen from false lumen, thereby assisting re-entry into the true lumen with a guidewire; and (2) facilitating true lumen entry of a CTO when a side branch arises just proximal to the CTO. Once a CTO is successfully crossed, the utility of IVUS involves more traditional applications that include precise vessel sizing and ensuring adequate stent expansion and apposition.

TRUE LUMEN RE-ENTRY METHOD
True lumen re-entry can be assisted with IVUS when the guidewire has passed into a subintimal space adjacent to the true lumen and attempts to re-enter the true lumen under fluoroscopic guidance are unsuccessful. In this situation, an IVUS catheter can be placed into the subintimal space after predilation with a 1.5-mm balloon catheter. The IVUS catheter can then be advanced into the subintimal tract and the true lumen visualized. A second guidewire can then be advanced alongside the IVUS catheter, and under IVUS visualization, the true lumen can be re-entered.

A limitation of this technique includes the fact that conventional IVUS provides only a thin cross-sectional view of the artery; therefore, the wire must re-enter the true lumen at a point directly adjacent to the imaging element. The need for subintimal dilation creates an unwanted larger false lumen, and the monorail design of existing IVUS catheters precludes wire exchanges. In addition, because of the physical attributes of current IVUS catheters, the distal end of the IVUS catheter must reside approximately 10.5 to 23 mm distal to the imaging element, requiring a relatively longer subintimal tract to be made (Figure 1).

SIDE BRANCH TECHNIQUE
This technique takes advantage of the fact that CTOs often arise just distal to a branch artery. In the side branch technique, the IVUS catheter can be placed into this side branch just proximal to the CTO, whereby the proximal cap and true lumen of the CTO can be visualized.5,6 A limitation of this technique is that a precise anatomic arrangement of a side branch (such as a diagonal coronary artery) must exist just proximal to a complete occlusion (such as the mid left anterior descending coronary artery). The angle of the side branch relative to the parent vessel and CTO must also be favorable for cross-sectional imaging catheters.

GENERAL PRINCIPLES OF FL-IVUS CATHETERS
The procedural ideal of any CTO intervention is to enter the true lumen at the proximal cap of the occlusion, stay intraluminal, and re-enter the true lumen after crossing the distal cap. FL-IVUS catheter technology is currently being developed for clinical use with these goals in mind. The main innovation with FL-IVUS is the ability to image in an antegrade (forward-looking) manner distal to the catheter tip. FL-IVUS catheters have previously been developed with an annular ring array of 64 elements at the tip of the catheter.7 In other designs targeted for clinical use, there is a single 45-MHz imaging transducer oriented at a 45° angle at the end of the catheter. The catheter rotates within a polymer sleeve and provides a forward-looking cone of visualization. This should result in visualization of the proximal cap and facilitate entry into the true lumen through the proximal cap of the CTO. Once within the CTO, the vessel walls can be visualized in front of the FL-IVUS catheter. By keeping the catheter centered in the vessel, the guidewire can be directed to maintain a true lumen position. The potential advantages of FL-IVUS technology are listed in Table 1.

THE PREVIEW CATHETER
The Preview catheter is currently undergoing preclinical and early clinical evaluation. It is a single-use, over-the-wire imaging catheter, and the distal imaging tip is shown in Figure 2. This catheter is advanced over a conventional 0.014-inch guidewire to the site of occlusion. It rotates at a rate of 3 to 5 revolutions per a second at a frame rate of 3 to 5 frames per second. The current generation of this catheter has a 45-MHz transducer at the tip and is compatible with a 7-F guide. The catheter provides a unique forward-looking view at a depth of 5 mm and allows visualization of the vessel borders distal to the tip of the catheter. Conceivably, the catheter can be advanced behind a guidewire into and through a CTO while maintaining a true lumen position. Figure 3 shows a schematic of the Preview catheter in clinical use.

FL-IVUS WITH RADIOFREQUENCY
A second-generation FL-IVUS catheter, which has not yet been named, is built on the Preview catheter platform, but also integrates a radiofrequency (RF) ablation element at the catheter tip. Therefore, in addition to imaging, it adds the elements of steering and tissue ablation. The tip of this catheter is angled, rotationally directable, and can be used to ablate CTO tissue under direct IVUS visualization; ablation time and power are user adjustable. Once the tissue is ablated, the catheter is gently advanced, and the vessel is reimaged. Sequential ablations can be directed to maintain an intraluminal position and avoid tracking toward the adventitia or subintimal space. The operation of the FL-IVUS catheter with RF is shown in Figure 4. This catheter has been used successfully in animal models of CTO, and early clinical evaluations are planned.

CONCLUSION
CTOs remain the most challenging subset of coronary percutaneous interventions with success rates averaging 50% in current practice. These procedures tend to be lengthy and require a high degree of operator skill, experience, and patience. The ultimate goal of any new CTO technology is to increase the procedural success rates of CTO crossing and to shorten the procedure time. FL-IVUS technology may provide an entirely new technique and approach to CTO intervention with direct visualization of the CTO distal to the catheter tip. Whether with guidewire-assisted or RF energy crossing, FL-IVUS has the potential to improve the ease and success of CTO interventions in the future.

Jason H. Rogers, MD, is from the Division of Cardiovascular Medicine, University of California Davis Medical Center in Sacramento, California. He has disclosed that he is a paid consultant to and receives grant/research funding from Volcano Corporation. Dr. Rogers may be reached at (916) 734-3764; jason.rogers@ucdmc.ucdavis.edu.

 

Contact Info

For advertising rates and opportunities, contact:
Craig McChesney
484-581-1816
cmcchesney@bmctoday.com

Stephen Hoerst
484-581-1817
shoerst@bmctoday.com

Charles Philip
484-581-1873
cphillip@bmctoday.com

About Cardiac Interventions Today

Cardiac Interventions Today (ISSN 2572-5955 print and ISSN 2572-5963 online) is a publication dedicated to providing comprehensive coverage of the latest developments in technology, techniques, clinical studies, and regulatory and reimbursement issues in the field of coronary and cardiac interventions. Cardiac Interventions Today premiered in March 2007 and each edition contains a variety of topics in a flexible format, including articles covering various perspectives on current clinical topics, in-depth interviews with expert physicians, overviews of available technologies, industry news, and insights into the issues affecting today's interventional cardiology practices.