Implantation Technique for Bioresorbable Scaffolds

Device description and patient/lesion selection with the Absorb BVS.

By Jiang Ming Fam, MD, MBBS; and Robert-Jan van Geuns, MD
 

The need for mechanical support for dilated vessels is temporary; beyond the first few months, there are potential disadvantages to the use of a permanent metallic prosthesis.1,2 Bioresorbable scaffolds (BRSs) were recently developed as an alternative to metallic stents. BRSs demonstrated complete bioresorption after approximately 3 years, accomplished by vessel lumen enlargement, a reduction of the plaque-to-media ratio, and restoration of vasomotion,3,4 essentially overcoming the limitations of metallic drug-eluting stents (DESs). The Absorb bioresorbable vascular scaffold (BVS; Abbott Vascular) is the most widely available commercial BRS and has demonstrated in early studies to be safe and efficacious when implanted in simple coronary lesions.5 Early studies have also suggested that implantation of the Absorb BVS is noninferior to second-generation DESs in terms of clinical outcomes, although these outcomes vary.6-8

These varying clinical outcomes have also been attributed to the lack of a standardized protocol in the implementation of a BRS-specific protocol, especially in light of a new device that is structurally different from the newer-generation DESs currently used. The Absorb BVS is constructed of a poly-L-lactide (PLLA) polymer backbone in a semicrystalline composition. To achieve sufficient radial strength, the current strut thickness is 156 μm, almost twice that of current metallic DESs. Even more, overexpansion of these semicrystalline scaffolds is limited to 0.5 mm to avoid rupture of the device during implantation.

Recently, there have been concerns regarding the risk of scaffold thrombosis (ST).9 Main pathomechanisms include factors pertaining to incomplete lesion coverage, underexpansion of the scaffold, and malapposition. The reported findings of a slightly higher thrombosis rate—early ST in particular—seem to be related to procedural factors such as the implantation technique necessary to compensate for the relative thick struts, similar to first-generation metallic bare-metal stents and DESs, and are therefore potentially preventable.10,11 The importance of a dedicated implantation protocol for BRSs, including preimplantation plaque modification, routine high-pressure scaffold postdilatation with noncompliant balloons, and the liberal use of intracoronary imaging, such as optical coherence tomography (OCT) to evaluate scaffold apposition and coverage, was highlighted in a recent study that showed that BRS thrombotic rates can be reduced by approximately 70% using a specific implantation technique.11 In this article, we discuss the standard implantation technique of the Absorb BVS and its potential use in more complex and challenging lesions.

DEVICE DESCRIPTION

The Absorb BVS is composed of a bioresorbable PLLA polymer backbone coated with a thin bioabsorbable poly-D,L-lactide, which allows everolimus to be eluted in a manner identical to that of the Xience V stent (Abbott Vascular). The scaffold is balloon-expandable, and other than two platinum radiopaque markers at each end of the scaffold, the rest of the scaffold is radiolucent, facilitating clear visualization on angiography and other imaging modalities such as cardiac CT. The scaffold edges on each end start within 1 mm of the balloon markers. The radiopaque scaffold markers are not located at the edges of the scaffold, but are placed approximately 1 mm from the proximal edge and 0.3 mm from the distal edge of the scaffold, which is worth noting, especially when implanting overlapping scaffolds.

The first generation of the Absorb BVS (revision 1.0; average strut thickness, 150 μm; crossing profile, 1.4 mm; circumferential out-of-phase zigzag hoops linked together by three longitudinal struts between each hoop) was tested in ABSORB Cohort A.5 This version unfortunately showed significant angiographic late loss and strut area reduction at 6 months, suggesting loss of radial support before vascular healing was completed.5 At 2-year follow-up, the concept of bioresorbable technology was finally demonstrated with resorption of the device, late lumen enlargement, restoration of vasomotion, and endothelial function.12,13 The second-generation Absorb BVS (revision 1.1) underwent modification in the strut design (in-phase zigzag hoops linked by bridges) and polymer manufacturing process. The metabolic process is slower in the second version, thus it provided more lasting mechanical support with reduced recoil. The polymeric scaffold maintains its radial strength for approximately 6 months after implantation and then undergoes a gradual metabolic process through the Krebs cycle into CO2 and H2O over the course of 2 to 4 years.14,15

GENERAL TECHNICAL CONSIDERATIONS

Technical considerations behind the successful implantation of the Absorb BVS are largely related to mechanical and structural properties of the polymeric scaffold structure. The current scaffold has relatively thick struts (156 µm) to maintain its radial strength. After the crimping process, the crossing profile (1.4 mm) is significantly larger than a contemporary metallic DES.16 This has implications on device delivery and trackability, creation of flow disturbances, and delays in scaffold re-endothelialization, possibly accounting for the differences seen in procedure duration, device success rates, and event rates in myocardial infarction (MI) and ST at 12 months.17-19 Table 1 summarizes the limitations and their impact on percutaneous coronary intervention (PCI).20 The situation is exacerbated by operators’ varying experience with Absorb BVS implantation, inconsistencies in device sizing, lesion preparation, routine high-pressure postdilation, and guidance with intracoronary imaging. Only 66.2% of patients treated with Absorb in four randomized trials had postdilation, and 23.9% underwent intracoronary imaging.21

A BRS-specific implantation protocol is based on following the five simple rules:

  • Prepare the lesion
  • Properly size the vessel
  • Pay attention to the expansion limits of the scaffold
  • Postdilate the BRS with a properly sized noncompliant balloon
  • Pay attention to dual antiplatelet therapy and patient compliance

PATIENT AND LESION SELECTION

In general, patients who may benefit from BRS implantation are younger and have longer lesions, as long metallic stents are associated with an increased risk of stent failure.22 Lesions that cannot be adequately predilated, such as when the predilation balloon cannot fully expand or when residual stenosis exceeds 40%, may not be considered for a BRS because the BRS with relatively thick struts would be underexpanded, increasing the potential risk of ST.11 Other factors to be considered include lesions that may pose potential deliverability issues, such as vessel tortuosity (vessels with extreme angulation of the segment proximal to the lesion) or heavily calcified lesions, and should be avoided. In such lesions, the thick polymeric struts of the scaffold with a large crossing profile may cause a buildup of friction between the device and lesion/catheter, increasing the risk of scaffold dislodgement during forceful movement in delivering the scaffold, especially when used by relatively inexperienced operators.21 Ideally, new operators of BRSs should start with stable, simpler lesions in stable patients and build up their expertise gradually before attempting implantation of the BRS in more complex lesions.

Rule 1: Prepare the Lesion

It is highly recommended to achieve adequate lesion preparation by using semi- or noncompliant balloons with a diameter equal or just undersized compared to the reference diameter of the BRS device selected (1:1 predilation). Sometimes, short, high-pressure balloons may be used to treat isolated segments of underexpansion.

Rule 2: Properly Size the Vessel

Exact vessel sizing and compliance to manufacturers’ guidelines on scaffold matching of lumen dimension are crucial. An analysis of the nominal BVS scaffold size to quantitative coronary angiography (QCA) maximum reference vessel diameter and to clinical outcomes was reported in a pooled patient-level analysis involving > 1,200 patients from three ABSORB studies.23 Subjects in the “scaffold oversize” group (defined as subjects with both proximal and distal maximum reference vessel diameters smaller than nominal scaffold size) experienced higher rates of major cardiovascular events and target vessel MI than those in the “scaffold non-oversize” group (defined as those in whom the proximal or distal maximum reference vessel diameter was larger than that of the scaffold). In ABSORB III, if vessels < 2.25 mm were excluded from the analysis, the incidence of ST was equivalent to the Xience EES.6 Compliance to vessel sizing guidelines may further improve target lesion failure by reducing the incidence of MI and ST. Thus, implanting BRSs in vessel sizes that are too small (mean reference vessel external elastic lamina diameter < 2.5 mm, which corresponds to a reference vessel diameter of 2.25 mm on QCA) should be avoided.

Because the Absorb BVSs have thicker struts, preferably, long regions of scaffold overlap should be avoided because it increases the risk of ST and side branch occlusion. In addition, polymers are invisible under x-ray, with the exception of two radiopaque edge markers. Therefore, placement of the scaffold can be difficult, especially in regions of significant overlap or foreshortening.

Rule 3: Pay Attention to the Expansion Limits of the Scaffold

Compared to DESs, BRSs have a limited range of expansion (ie, 0.5 mm more than the reference diameter) due to their polymeric composition, limiting their use in cases of vessel tapering (Table 2). Huge malapposition can be uncorrectable and persist at follow-up until resorption occurs, and attempts to correct large malapposition by overexpansion with a large balloon can lead to scaffold disruption (Figure 1).

Figure 1. Malapposed struts (indicated by *) (A). Attempts to correct large malapposition by overexpansion with a large balloon can lead to acute disruption of the scaffold (disrupted struts indicated by +) (B).

Scaffold implantation. Delivery of the BRS requires the application of gentle constant pressure to the lesion through a 6 F or larger guiding system. The main difference from the metallic stents is that balloon inflation for scaffold deployment should be gradual (2 atm every 5 seconds), with a minimum inflation duration of 30 seconds. When difficulties are encountered during delivery of the BRS, active support with the guiding catheter in the form of deep-vessel intubation may alleviate this problem, but there is a risk of coronary dissection with potentially disastrous consequences. Other methods to overcome this problem include vessel straightening with a second buddy wire24 or buddy balloon,25 the use of an anchor balloon,26 and the use of an extra back-up support guiding catheter or a guide extension.27

Rule 4: Postdilate the BVS With a Properly Sized Noncompliant Balloon

Operators should aim to achieve to cover ≥ 2 mm of the healthy vessel at either edge of the treated lesion using the BRS, with a result of < 10% residual stenosis after deployment of the BRS with optimal scaffold expansion and apposition.1 Performing postdilation using high-pressure balloons (> 16 atm) for 10 to 30 seconds is frequently required, and an optimal result is confirmed using intravascular imaging.

Role of intravascular imaging. The application of IVUS and OCT, a light-based intravascular imaging technology,28 has enabled us to address the challenges facing the use of BRSs and overcome the limitations of angiography in assessing the significance of coronary stenosis and the results of PCI in a clinical setting.29 The advantages of using intracoronary imaging include the capability to provide accurate luminal measurements and optimal detection of scaffold malapposition and fracture, which cannot be reliably detected on a simple angiogram. One main advantage of IVUS is the high depth of penetration that allows direct and easy visualization of total vessel diameter and area, allowing operators to optimize scaffold size without increasing the risk of disruption with oversized balloons.30

Figure 2. “Virtual PCI” planning. OCT images of the left anterior descending artery before scaffold implantation showing the distal landing zone (A), minimal lumen area (B), and proximal landing zone (C). OCT can assess lumen dimensions accurately, assess underlying plaque composition, and shows the location of the minimal lumen area in relation to the treated vessel on the lumen profile (D). In this way, OCT can guide scaffold implantation strategy by assessing the scaffold length (18 mm, in this example) required for optimal lesion coverage and avoiding SB ostia. The SB is seen on the longitudinal profile (E). The LAD is shown before (F) and after (G) implantation of a 3- X 18-mm Absorb BVS (dashed line). Aft, guidewire artifact.

There are also advantages to using OCT in the deployment of BRSs. OCT was one of the most useful techniques for the early evaluation of the Absorb BVS and its resorption process. New OCT probes are low-profile (2.6–2.7 F), flexible, coated with a hydrophilic layer, and the acquisition speed is at least 10 times higher when compared to IVUS. OCT catheters have a low delivery profile and can pass almost every lesion with few anatomical or patient exclusion criteria. The OCT imaging procedure is safe30 and fast, providing all necessary information in just seconds (Figure 2).31 The latest European Society of Cardiology guidelines on myocardial revascularization has already recommended OCT as a tool in selected patients to optimize stent implantation (level of evidence class II B, level C).29

An improvement in clinical and angiographic outcomes with intracoronary imaging-guided PCI was first shown in the use of IVUS.32 In a meta-analysis, IVUS-guided DES deployment compared with standard angiographic guidance was associated with a reduced incidence of major adverse cardiac events.33 The advantages of imaging guidance observed by IVUS may also apply to OCT, which was supported by findings of an observational study in which angiographic plus OCT guidance was associated with a significantly lower risk of cardiac mortality or MI, even after multivariate adjustment or propensity score-adjusted analyses.34 Preliminary observations support a potentially beneficial role of OCT during BVS implantation in improving outcomes.9

Rule 5: Pay Attention to Dual Antiplatelet Therapy and Patient Compliance

A minimum duration of 12 months of dual antiplatelet therapy is suggested in both stable and acute patients treated with a BRS. However, an even longer duration of dual antiplatelet therapy (18–24 months) and/or more potent antiplatelet medications, such as ticagrelor or prasugrel, may be used depending on the risk-benefit balance between thrombotic/ischemic risk and bleeding risk.

CONCLUSION

Procedure- and lesion-related factors play an important role in acute procedural success, and various technical difficulties can be encountered in the implantation of the Absorb BVS. The impact of these factors may be mitigated by systematically applying the BRS implantation recommendations. The long-term success of BRSs relies on a combination of careful patient and lesion selection with proper implantation technique and optimization of procedural results in the clinical arena. By taking proactive steps to perform accurate patient/lesion selection and implement an optimal implantation technique, risks of adverse outcomes will be minimized, translating into improved clinical outcomes.

1. Tamburino C, Latib A, van Geuns RJ, et al. Contemporary practice and technical aspects in coronary intervention with bioresorbable scaffolds: a European perspective. EuroIntervention. 2015;11:45-52.

2. Stefanini GG, Holmes DR Jr. Drug-eluting coronary-artery stents. N Engl J Med. 2013;368:254-265.

3. Simsek C, Karanasos A, Magro M, et al. Long-term invasive follow-up of the everolimus-eluting bioresorbable vascular scaffold: five-year results of multiple invasive imaging modalities. EuroIntervention. 2016;11:996-1003.

4. Karanasos A, Simsek C, Gnanadesigan M, et al. OCT assessment of the long-term vascular healing response 5 years after everolimus-eluting bioresorbable vascular scaffold. J Am Coll Cardiol. 2014;64:2343-2356.

5. Ormiston JA, Serruys PW, Regar E, et al. A bioabsorbable everolimus-eluting coronary stent system for patients with single de-novo coronary artery lesions (ABSORB): a prospective open-label trial. Lancet. 2008;371:899-907.

6. Ellis SG, Kereiakes DJ, Metzger DC, et al. Everolimus-eluting bioresorbable scaffolds for coronary artery disease. N Engl J Med. 2015;373:1905-1915.

7. Gao R, Yang Y, Han Y, et al. Bioresorbable vascular scaffolds versus metallic stents in patients with coronary artery disease: ABSORB China trial. J Am Coll Cardiol. 2015;66:2298-2309.

8. Kimura T, Kozuma K, Tanabe K, et al. A randomized trial evaluating everolimus-eluting Absorb bioresorbable scaffolds vs. everolimus-eluting metallic stents in patients with coronary artery disease: ABSORB Japan. Eur Heart J. 2015;36:3332-3342.

9. Karanasos A, Van Mieghem N, van Ditzhuijzen N, et al. Angiographic and optical coherence tomography insights into bioresorbable scaffold thrombosis: single-center experience. Circ Cardiovasc Interv. 2015;8:pii:e002369.

10. Cassese S, Byrne RA, Ndrepepa G, et al. Everolimus-eluting bioresorbable vascular scaffolds versus everolimus-eluting metallic stents: a meta-analysis of randomised controlled trials. Lancet. 2016;387:537-544.

11. Puricel S, Cuculi F, Weissner M, et al. Bioresorbable coronary scaffold thrombosis: multicenter comprehensive analysis of clinical presentation, mechanisms, and predictors. J Am Coll Cardiol. 2016;67:921-931.

12. Serruys PW, Ormiston JA, Onuma Y, et al. A bioabsorbable everolimus-eluting coronary stent system (ABSORB): 2-year outcomes and results from multiple imaging methods. Lancet. 2009;373:897-910.

13. Onuma Y, Serruys PW, Gomez J, et al. Comparison of in vivo acute stent recoil between the bioresorbable everolimus-eluting coronary scaffolds (revision 1.0 and 1.1) and the metallic everolimus-eluting stent. Catheter Cardiovasc Interv. 2011;78:3-12.

14. Oberhauser JP, Hossainy S, Rapoza RJ. Design principles and performance of bioresorbable polymeric vascular scaffolds. EuroIntervention. 2009;5(suppl F):F15-F22.

15. Serruys PW, Onuma Y, Garcia-Garcia HM, et al. Dynamics of vessel wall changes following the implantation of the absorb everolimus-eluting bioresorbable vascular scaffold: a multi-imaging modality study at 6, 12, 24 and 36 months. EuroIntervention. 2014;9:1271-1284.

16. Iqbal J, Onuma Y, Ormiston J, et al. Bioresorbable scaffolds: rationale, current status, challenges, and future. Eur Heart J. 2014;35:765-776.

17. Stone GW, Gao R, Kimura T, et al. 1-year outcomes with the Absorb bioresorbable scaffold in patients with coronary artery disease: a patient-level, pooled meta-analysis. Lancet 2016;387:1277-1289.

18. Capodanno D, Gori T, Nef H, et al. Percutaneous coronary intervention with everolimus-eluting bioresorbable vascular scaffolds in routine clinical practice: early and midterm outcomes from the European multicentre GHOST-EU registry. EuroIntervention 2015;10:1144-53.

19. Panoulas VF, Miyazaki T, Sato K, et al. Procedural outcomes of patients with calcified lesions treated with bioresorbable vascular scaffolds. EuroIntervention. 2016;11:1355-1362.

20. Rizik DG, Hermiller JB, Kereiakes DJ. The ABSORB bioresorbable vascular scaffold: a novel, fully resorbable drug-eluting stent: current concepts and overview of clinical evidence. Catheter Cardiovasc Interv. 2015;86:664–677.

21. Ishibashi Y, Onuma Y, Muramatsu T, et al. Lessons learned from acute and late scaffold failures in the ABSORB EXTEND trial. EuroIntervention. 2014;10:449-457.

22. Kobayashi Y, De Gregorio J, Kobayashi N, et al. Stented segment length as an independent predictor of restenosis. J Am Coll Cardiol. 1999;34:651-659.

23. Ishibashi Y, Nakatani S, Sotomi Y, et al. Relation between bioresorbable scaffold sizing using QCA-Dmax and clinical outcomes at 1 year in 1,232 patients from 3 study cohorts (ABSORB Cohort B, ABSORB EXTEND, and ABSORB II). JACC Cardiovasc Interv. 2015;8:1715-1726.

24. Burzotta F, Trani C, Mazzari MA, et al. Use of a second buddy wire during percutaneous coronary interventions: A simple solution for some challenging situations. J Invasive Cardiol. 2005;17:171-174.

25. Li SS, Cheng CW. Coronary angioplasty on an impassable calcified stenosis using a buddy balloon technique. Catheter Cardiovasc Interv. 2004;62:35-37.

26. Fujita S, Tamai H, Kyo E, et al. New technique for superior guiding catheter support during advancement of a balloon in coronary angioplasty: the anchor technique. Catheter Cardiovasc Interv. 2003;59:482-488.

27. Rao U, Gorog DA, Szygula J, et al. The GuideLiner “child” catheter. EuroIntervention. 2010;6:277-279.

28. Lowe HC, Narula J, Fujimoto JG, Jang IK. Intracoronary optical diagnostics current status, limitations, and potential. JACC Cardiovasc Interv. 2011;4:1257-1270.

29. Kolh P, Windecker S. ESC/EACTS myocardial revascularization guidelines 2014. Eur Heart J. 2014;35:3235-3236.

30. Roy P, Steinberg DH, Sushinsky SJ, et al. The potential clinical utility of intravascular ultrasound guidance in patients undergoing percutaneous coronary intervention with drug-eluting stents. Eur Heart J. 2008;29:1851-1857.

31. van der Sijde JN, Karanasos A, van Ditzhuijzen NS. Safety of optical coherence tomography in daily practice: a comparison with intravascular ultrasound [published online March 18, 2016]. Eur Heart J Cardiovasc Imaging.

32. Foin N, Lee R, Mattesini A, et al. Bioabsorbable vascular scaffold overexpansion: insights from in vitro post-expansion experiments. EuroIntervention. 2016;11:1389-1399.

33. Shin DH, Hong SJ, Mintz GS, et al. Effects of intravascular ultrasound-guided versus angiography-guided new-generation drug-eluting stent implantation: meta-analysis with individual patient-level data from 2,345 randomized patients. JACC Cardiovasc Interv. 2016;9:2232-2239.

34. Prati F, Di Vito L, Biondi-Zocca G, et al. Angiography alone versus angiography plus optical coherence tomography to guide decision-making during percutaneous coronary intervention: the Centro per la Lotta contro l’Infarto-Optimisation of Percutaneous Coronary Intervention (CLI-OPCI) study. EuroIntervention 2012;8:823-829.

Jiang Ming Fam, MD, MBBS
Consultant
Department of Cardiology
National Heart Centre Singapore
Singapore
Disclosures: None.

Robert-Jan van Geuns, MD, PhD, FESC, FACC
Professor of Interventional Cardiology
Department of Cardiology
Thorax Centre
Erasmus Medical Centre
Rotterdam, The Netherlands
r.vangeuns@erasmusmc.nl
Disclosures: Research grants and speaker fees from Abbott Vascular.

 

Contact Info

For advertising rates and opportunities, contact:
Craig McChesney
484-581-1816
cmcchesney@bmctoday.com

Stephen Hoerst
484-581-1817
shoerst@bmctoday.com

Charles Philip
484-581-1873
cphillip@bmctoday.com

About Cardiac Interventions Today

Cardiac Interventions Today is a publication dedicated to providing comprehensive coverage of the latest developments in technology, techniques, clinical studies, and regulatory and reimbursement issues in the field of coronary and cardiac interventions. Cardiac Interventions Today premiered in March 2007, with its launch issue focusing on the state of coronary drug-eluting stent use. Each edition will cover a variety of topics in a flexible format that includes articles covering various perspectives on current clinical topics, in-depth interviews with expert physicians, overviews of available technologies, industry news, and insights into the issues affecting today's cardiology practices.